
MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 12 of 20
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
9 - 12
Image Quality
7.
Select the Procedure
Calibration
from the examination tab card.
8.
Set the following exposure parameter on the control console of the MAMMOMAT
Novation
DR
:
Mo/Mo, 28KV, 250mAs
9.
Insert the collimator mounted plexi phantom.
10.
Mount the compression plate simulator.
11.
Double click on the
first
Image of the 8 at the
Procedure: Calibration Procedure
12.
Start an exposure.
Wait until the image is loaded into the viewer. The following lines are displayed in
the message bar.
Image acquisition is in progress .....
Saving processed Image .....
Loading Image ......
13.
Check the image quality of the image. No objects or artifacts shall be visible in the
image.
If you
Accept
the image, the image will be used for Gain Calibration.
Remember, you can always add or remove images to
the "Procedure" by using the right mouse click.
Fig. 13 Start Calibration
NOTE