Cable connections
5 - 11
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 11 of 14
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Connecting the incoming mains to the stand (400 V, 2-phase)
5
400 V 2-phase connection
5
1.
Connect the incoming mains cable (1) to X899 (2) in the stand as shown in Fig. 13
and Fig. 14.
2.
Fit plastic tubes (4) on the plates.
3.
Strain-relieve (3) the incoming mains cable.
Fig. 13 Incoming mains cable
Fig. 14 Incoming mains, 2-phase
1
MAM00097
L1
L2
L3
N
1
2
3
1
4
1
TWO-PHASE
GENERATOR
STAND
F10
F20
X881
Incoming mains
208, 230, 240, 277 or 400 VAC
P.E N
L1 L2 L3
X899
MA
M00
095