Connecting the service PC
20
The service PC must be connected with connecting cable part no. 99 00 440 RE999 to
p.c. board in the generator (do not insert the diskette in the drive yet). Use port COM 1 on
the service PC.
Starting up and using the service PC
20
1
Switch on generator and service PC.
After initialization, the service PC shows: C:\>
2
Now insert the discette with the service program.
3
Select the appropriate drive
and then press <ENTER>.
The screen shows: A:\> or B:\>
4
Start the service program by typing ’SERVICE’, then press <ENTER>.
The program asks for the user’s name:
Your name, please
Type the name of the technician, for example NN, and then press <ENTER>.
5
The program asks for the password:
Password, please
Type the password (
∗∗∗∗∗∗∗∗
) and then press <ENTER>.
The display window shows:
Main menu
6
Select the program part to be used with keys <
↑
> and/or <
↓
>, then press <ENTER>.
The program part selected is shown with a background:
If necessary, additional subroutines can be similarly selected here.
7
Make the necessary entries in the appropriate part of the program.
Save the entered data with <F2>. Page back in the program with <ESC>.
The appropriate instructions are shown on the monitor.
8
End the procedure with the service PC with <F10>.
XXX - - XXX - -
XXX - - XXX
By this means, the way is shown how to call up a particular sub-
routine (display windows).
For example:
Main menu
Configuration
Iontomat
Corr. curve
{
A:}
{B:}
Configuration