Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 6 of 10
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
18 - 6
Final procedures
Mounting the front cover onto the generator
18
1.
Fit the front cover (3) to the generator and fasten it with sixteen screws (2).
2.
Press on the outer cover strip (1).
Final protective grounding resistance test
18
Using the protective ground wire tester, measure the resistance between the mains
grounding connection (incl mains cable) and any metallic part of the stand and generator
which the patient/operator may come in contact with.
The protective ground resistance must not exceed 0.2
Ω
.
Fig. 5
Mounting the front cover
When attaching the front cover, make sure that all eight contact
washers (four on either side) are used again as in the original
assembly (under the screws with clearance holes in the plastic
strips). The screws must be securely tightened and the washers
must be functional. This is necessary to establish protective
ground connection.
3
1
2
MA
M00
204
WARNING