Cable connections
5 - 7
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 7 of 14
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Rotating anode cable X9
5
1.
Connect the rotating anode cable X9 to D711 X9 (1).
2.
Connect the two blue free wires from connector D711 X9 (1) to capacitor C51 (2).
3.
Clamp the cable to the chassis with a cable clamp (3).
Fig. 7
Connecting X9 to the generator
When fitting the cable clamp, ensure satisfactory contact
between shield braid (4) and chassis.
2
1
MAM00163
4
3
CAUTION