15 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 4
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Checking the compression device and OPCOMP
15
Checking the compression device
15
1.
Attach compression plate (1/Fig. 1) if not already done.
2.
Check that the compression plate cannot be removed without pressing the
compression plate release button (2/Fig. 1).
3.
Check the function of the knobs for manual compression/decompression (3/Fig.
1).
4.
Place a piece of soft material on the object table, for example a rolled up towel or a
piece of foam-rubber (4/Fig. 1).
5.
Check the function of the foot switch for motorized compression/decompression
(1) and (2) in (Fig. 2).
Fig. 1
Swivel-arm system
Fig. 2
Foot switch
1
2
3
4
5
MA
M001
99
MAM00132
1
2