SP B7-230.033.07
Page 2 of 2
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
12 - 2
Further programming
Reducing the generator power
12
To achieve full power, the measured resistance must not exceed the following values:
•
0.25
O
at 110 V (1-phase)
•
0.45
O
at 208 V
•
0.50
O
at 230 V
•
0.60
O
at 240 V
•
0.65
O
at 277 V
•
0.85
O
at 400 V (2-phase)
If the above values are exceeded, reduce the generator power as follows:
Main menu:
12
1.
Select Configuration <
↑
> <
↓
> < ENTER>.
2.
Select Miscellaneous <
↑
> <
↓
> < ENTER>.
3.
Select Red. of power <
↑
> <
↓
> < ENTER>.
Reduction of power depending on line quality.
Tube is P 40 MoW.
4.
Select nominal line voltage ____ V with cursor keys <
↑
> <
↓
> and < ENTER>.
5.
Type in line impedance ____
Ω
. Enter the measured line impedance.
Calculated power is ___ kW. Max. possible tube power is displayed.
6.
Save the programmed values with <F2>.