16 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 2
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Checking the exposure blocking
16
Checking the exposure blocking
16
Check that the exposure release is blocked during the following conditions:
•
No film cassette inserted.
•
No object table installed.
Fig. 1
Swivel-arm system
The "casette loaded" switch must be set to "ON" with the service
PC program: Configuration - Miscellaneous - Cassette loaded
check.
The film cassette must be exchanged after each exposure. A sec-
ond exposure before exchanging the casette is not possible
(exposure release is blocked).
MA
M0
0200
NOTICE
NOTICE