Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 2 of 6
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
4 - 2
Installing the generator and the stand
Free-standing radiation shield (option)
4
For the installation of the free-standing radiation shield, please refer to separate instruc-
tions enclosed with the radiation shield.
Removing the generator cover
4
1.
Pry loose the plastic strip (1) and remove the sixteen screws (2), eight on each
side.
2.
Remove the front cover (3) from the generator.
Pay special attention to the contact washers (there are four con-
tact washers on either side). They are needed again when reas-
sembling the front cover to establish protective ground
connection and to fulfil EMC (Electro Magnetic Compatibility)
requirements.
Fig. 2
Removing the generator cover
NOTICE
3
1
2
M
A
M
0009
1