Cable connections
5 - 9
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 9 of 14
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
3.
Shield the control console cable to the bottom plate of the generator as shown in
Fig. 10.
4.
Provide the cables with ferrite sleeves. Cables X5 and X32 can be laid together in
one ferrite sleeve, X30 and the decompression-button cable in another sleeve
(two extra ferrite sleeves are included in the delivery). Plug together the decom-
pression button cables. Shield the decompression button cable to the bottom plate
, see Fig. 10. Lay the decompression button cable in the cable duct, alter-natively
underfloor, to the stand, see Fig. 4 in Chapter 4 "Installing the generator and the
stand".
5.
Connect X5 and X32 to the corresponding sockets on circuit board D700, X30 to
connector T3/X30 at the side of the circuit board chassis. Connect the ground wire
to the protective ground terminal below fuse F10, see Fig. 11.
Fig. 10 Cables from separate control console
Fig. 11 Connecting the cables from the separate control console
Ground wire
Decompression
button cable
X5, X32
X30
MA
M0
0174
F10
F20
Protective
ground
F
3
F4
F5
X32
X5
T3/X30
MAM00164