Values recorded during installation of Mammomat:
Keep measuring protocol in technical documentation.
Tube P40 MoW, tungsten anode 0.3 (MAMMOMAT 3000 only)
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5 %
______V x ____ s x 40 mA/V = ____ mAs
20 mAs
&
10 %
Date:______________
Sign:______________
Tube P40 MoW, molybdenum anode 0.3
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5 %
______V x ____ s x 40 mA/V = ____ mAs
20 mAs
&
10 %
Date:______________
Sign:______________
Tube P40 MoW, tungsten anode 0.15 (MAMMOMAT 3000 only)
Calculated values
Required values
______V x 5 kV/V = _____kV
30 kV
&
5 %
______V x ____ s x 40 mA/V = ____ mAs
10 mAs
&
10 %
Date:______________