Operating Instructions
Chapter
4
CLX 490
8 009 237/0000/07-05-2002
©
SICK AG · Division Auto Ident · Germany · All rights reserved
4-9
Installation
Parameterizing an external sensor as a trigger source:
These settings are not required if the CLX is operated with the default configuration.
1.
Choose the D
EVICE
C
ONFIGURATION
tab.
2.
In the section R
EADING
T
RIGGER
M
ODE
click the option S
ENSOR
I
NPUT
(A
CTIV
H
IGH
).
3.
Download the settings to the CLX by clicking
in the toolbar.
The D
OWNLOAD
P
ARAMETERS
dialog box is then displayed.
4.
Confirm the dialog box by choosing P
ERMANENT
.
The CLX operates with the "Sensor" switching input as an external trigger source. The
reading pulse starts when the input is energized (high).
4.4.3
Mounting the sensors for detecting the object distance
If the dynamic focus control function of the CLX is triggered by external sensors, suitable
sensors must be mounted in the vicinity of the CLX.
operation from above. The sensors must be arranged in such a way that all of the potential
object heights are classified uniquely and overlapping reading ranges occur that can be
formed with the depths of field of the CLX. A maximum of five switching inputs are available
for this purpose. A max. of 32 switching states for 8 distance configurations (reading ranges)
can be implemented using the internal assignment table (combinations). Photoelectric reflex
switches, for example, can be used to detect the object distance. In the default setting, all
five "IN 0... IN 4" switching inputs are selected for dynamic focus control.
Fig. 4-7:
Mounting example for object distance detection
;
;
;
LS1
LS2
LS7
h2
h7
h1
Reading distance
DC = Distance Configuration
Object distance
DC 8
DC 2
DC 1