Figure 34: Alignment of device flange
1
Optical alignment, horizontal
2
Optical alignment, vertical
•
Horizontal adjustment: Horizontal
adjustment of the focus
•
Vertical adjustment: Vertical adjust‐
ment of the focus.
3
Check the values for the light energy:
–
The values for the light energy V1 ... V4 must be in the range 250 ... 500.
–
the values must approximately have the same size.
b
If you cannot see the focal point or adjustment is not possible, check:
–
Correct distance between device flange and purge air fixture (
"Installing the device flange on the purge air fixture", page 53
–
see "Installing the device flange on the purge air
–
Contaminated windows,
see "Cleaning the window", page 77
–
Defective sender lamp,
see "Replacing the sender lamp",
NOTE
Recommendation: Start check cycle after each manual alignment and maintenance
work.
3
The values for the light energy V1 .. V4 must be in the range 250 ... 500 and
approximately have the same size.
NOTE
If you cannot see a focal point or adjustment is not possible, check
b
that the gap between the device flange and the purge air fixture is correctly
adjusted,
see "Installing the device flange on the purge air fixture", page 53
b
GMP probe: Whether the lever is set to "open",
see "Installing the SR-unit on
.
b
for dust or humidity in the gas duct.
b
contamination of windows,
see "Cleaning the window", page 77
.
b
function of the sender lamp,
see "Replacing the sender lamp", page 80
NOTE
Empfehlung: Kontrollzyklus nach jeder manuellen Ausrichtung und Instandhal‐
tungsmaßnahme starten.