8.3.4.1
Diagnosis
The “Diagnosis” menu shows the current error messages.
the following information is available for troubleshooting:
•
The GM32 signals a malfunction or an unsafe operational state with status signals
(option). See electrical connection diagram.
•
Logbook: Access to the logbook via SOPAS ET,
.
•
List of error messages and measures for clearing malfunctions,
Status messages, status indicators and system states
Status
LED
Significance
Measured
value dis‐
play
Analog out‐
puts
1
Status sig‐
nal
2,3
Power on
(power volt‐
age on)
Power supply in order.
--
Uncertain
Measured
value
blinks
Measured value uncertain (e.g. outside calibration
range)
Cause: Press the DIAG button
All messages in the SOPAS ET logbook.
Clearing malfunctions,
Current
According to
setting
Maintenance
request
(mainte‐
nance
request mes‐
sage)
•
Irregularities (e.g. gas temperature too high, devi‐
ation from check cycle too high) that require a
review of the cause.
•
The measured values are valid.
•
Press the DIAG button: Error display
•
All messages in the SOPAS ET logbook
•
Clearing malfunctions,
Failure
•
Device failure (e.g. lamp failed).
•
Cause: Press the DIAG button.
•
All messages in the SOPAS ET logbook.
•
Clearing malfunctions,
Last valid measured value
held
1
Option
2
Option. Refer to the delivered System Documentation for the assignment of the status outputs.
3
See SOPAS ET in the “Digital outputs” menu.
8.3.4.2
Check cycle - results of last check cycle
NOTE
The results of the last check cycle can be called up in this menu item.
8
OPERATION
68