3.2.2
Device variants
“Basic” variant
•
Reference cycle,
see "Reference cycle", page 18
: Correction of internal drifts.
Zero point check.
•
Automatic mirror tracking: Automatic adjustment of optical axis.
•
Logbook: System messages are recorded in a logbook.
•
Network: Ethernet interface with OPC standard.
•
Ethernet: 1 x fiber optics connection on connection unit.
“Pro” variant
As “Basic” variant. In addition:
•
TÜV-tested for equipment subject to authorization:.
•
Check cycle,
: Reference cycle (as in the “Basic” var‐
iant) followed by a cycle to check and output the zero and check point. The check
cycle creates the QAL3 values (quality assurance of automated measuring sys‐
tems). The QAL3 values can be displayed with SOPAS ET.
•
Operator panel: Measured values, operational state and malfunction messages
are displayed in clear text on a monitor.
•
QAL3 Tool (CUSUM chart).
3.2.3
Options
•
I/O modules (Analog Out, Digital Out, Digital In, Analog In).
•
SCU: Operating unit to control several analyzers with SCU capability (see SCU
Operating Instructions).
•
Super calibration: Several applications/calibrations, e.g., for spare devices.
•
Measuring range switch-over (analog outputs). The valid measuring range is sig‐
naled via a digital output (configured).
•
Weatherproof cover.
3.3
SOPAS ET (PC program)
SOPAS ET enables:
•
Additional configuration
•
Access to logbook GM32
SOPAS ET runs on an external PC connected via the Ethernet,
, to the GM32.
NOTE
Further information on SOPAS ET:
•
Technical Information GM32
•
SOPAS ET Help menu
PRODUCT DESCRIPTION
3
8015558/YYJ8/V1.7/2016-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
17
Subject to change without notice