Initiator
Text
Classifica‐
tion
Description
Possible causes/remedy
System
Lamp perform‐
ance
M
aint.
Warning: Lamp performance < 20%.
Prepare to exchange the lamp,
.
Lamp perform‐
ance limit
M
aint.
Lamp performance too low.
Lamp mini‐
mum
M
aint.
Lamp adjustment: Signal too high for minimum
lamp current and exposure.
Check parameter settings.
Lamp 4Q max
M
aint.
The lamp current must be set to 1000 mA (stop)
in the alignment procedure.
•
Alignment, check the optics,
ment check (option) - automatic optical
alignment check", page 69
.
•
Possible lamp change,
•
Correct parameter settings.
Flashcard
missing
M
aint.
No Flashcard found.
Insert the Flashcard, replace a possibly defec‐
tive card.
IO com.
M
aint.
Communication fault to IO block.
•
Connection interrupted, check the cable.
•
Defective CAN-bus interface.
Spectro no
answer
M
aint.
No data received from spectrometer.
Malfunction on the interface to the spectrome‐
ter. Check the plug.
Cycle span
drift
M
aint.
The measurement on the control filters showed
an excess deviation.
The reference from the adjustment is not cor‐
rect. Check limit value parameter setting.
Cycle zero drift
M
aint.
The zero point measurement of a measured
value showed an excess deviation.
Check limit value parameter setting.
Cycle wave‐
length drift
M
aint.
.
The check of the current Lambda_C0 coefficient
showed an excess deviation.
Cycle peak
position
M
aint.
The check of the position of the peak of the NO
cell showed an excess deviation
•
Check limit value parameter setting.
•
Defective NO cell.
Cycle peak
width
M
aint.
The check of the peak width of the NO cell
showed an excess deviation.
Cycle cell
empty
M
aint.
During the check of the NO cell it is determined
that the highest measured extinction value in
the evaluation range is <0.1.
Cell is empty.
Temp control
voltage low
M
aint.
The measured voltage supply value is too small
(< 20 V).
Malfunction of temperature control unit.
Temp control
lamp fan
M
aint.
The lamp fan has a malfunction.
Malfunction:
•
Temperature control unit
•
Fan
•
Cabling
Temp control
optic fan
M
aint.
The fan of the optic carrier has a malfunction.
Temp control
spectro fan
M
aint.
The fan of the spectrometer has a malfunction.
Temp control
electronic
temp
M
aint.
Temperature of the temperature control elec‐
tronics >100 °C.
Malfunction of temperature control unit.
Temp control
spectro temp
M
aint.
SR-unit is too warm or too cold.
•
In the heating phase: Normal.
•
When running: Check ambient tempera‐
ture.
Data logging:
writing data
M
aint.
Error when writing logging data to the Flashcard. Flashcard memory full, Flashcard defective.
Data logging:
open file
M
aint.
Error when opening a file for logging data on the
Flashcard.
System I/O
Error
M
aint.
Error in “Modular I/O System”.
Incorrect configuration of I/O module or I/O
module defective.
Table 12: Error table: Maintenance request
10
TROUBLESHOOTING
90
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
8015558/YYJ8/V1.7/2016-07 | SICK