8.3.4.4
Adjustment
NOTICE
b
Perform this work only when the SR-unit is at operating temperature (in operation
for at least 30 minutes).
MAINTENANCE
REQUEST
Measuring
Back
1
2
3
A: 858
B: 818
C: 760
D: 728
X: -0.011
Y: 0.016
Alignment
1.2
/Menü/Align
Figure 33: Shows the optical alignment on the device display
1
Light energy
2
Focal point
3
X/Y values
Performing the optical alignment
1
Press the
Start
button. A crosshair with a focal point and X/Y values are shown on
the screen.
NOTE
The focal point is in the center of the crosshair when the following tolerances are
shown:
•
X: -0.05 ... +0.05
•
Y: -0.05 ... +0.05
2
Perform the optical alignment: Adjust the optical alignment by turning both adjust‐
ment screws on the device flange of the SR-unit (19 mm wrench).
NOTICE
The display on the monitor reacts to the adjustment with a delay.
b
Perform adjustments slowly and wait for approx. 20 seconds until the display
on the monitor has been updated.
8
OPERATION
70
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
8015558/YYJ8/V1.7/2016-07 | SICK
Subject to change without notice