7.9
Installing the SR-unit on the device flange
CAUTION
Damage when device is dropped
Its weight and size make the SR-unit with installed probe hard to handle. This increases
the risk that the measuring device is damaged when dropped.
b
Install the SR-unit and the probe separately.
1
2
3
4
5
Figure 27: Hinge of sender unit
1
Hinge
2
Hinge pin
3
Device flange
4
Window
5
Desiccant cartridges
1
Fit the SR-unit on the device flange into the hinge.
NOTE
Recommendation: Install the flange on the left.
The device flange can be installed on the left or right. The advantage of installa‐
tion on the left is the unhindered access to the device display when working on the
swiveled out flange.
2
Insert the hinge pin from above in the hinge.
NOTICE
If the hinge pin has not been correctly inserted, the SR-unit can drop when swiv‐
eled out.
Ensure the hinge pin is fully inserted.
3
Check the window for cleanness and clean, if required,
.
4
Check the desiccant cartridge is dry,
see "Check/replace the desiccant cartridge",
.
5
Close the SR-unit with the 4 quick-release fasteners.
6
GMP probe: Set the lever on the purge air fixture to the “open” position.
COMMISSIONING
7
8015558/YYJ8/V1.7/2016-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM32-Ex
59