START-UP AND PARAMETER SETTINGS
4
4.4.2
Factory settings
The steps required to modify these settings are described in the following Sections. For this
purpose, the devices must be connected in SOPAS ET (
see “Connection to the device via
), the Level 1 password set and the “Maintenance” mode set.
Parameter
Value
Function check
Every 8 h; output of control values (every
90 s) on standard analog output
Analog output (AO)
[mA]
Live zero (LZ)
4
Upper measuring range value
(MBE)
20
Current during maintenance
0.5
Current by malfunction
21 (optional 1)
Response time
60 s for all measured variables
Measured variable
Output on AO
Value at LZ
Value at MBE
Opacity [%]
1
0
100
Dust concentration a.c. ext [mg/m³]
2
0 200
Dust concentration a.c. SI [mg/m³]
3
0
200
Coefficients set (for dust concentration ext)
0.00 / 1.00 / 0.00
Coefficients set (for dust concentration SI)
0.00 / 1.00 / 0.00