57
8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
START-UP AND PARAMETER SETTINGS
4
4.1.4
Connection to the device via Ethernet (option)
Recommended procedure:
1
The MCU must be switched off.
2
Connect the MCU with the network.
3
Connect the PC to the same network.
4
Switch the MCU on.
5
Start SOPAS ET.
6
“Search settings”
7
“Device family oriented search”
8
Click the desired MCU
9
Make the settings:
– Ethernet communication (always clicked)
– USB communication (always clicked)
– Serial communication: Do
not
click
10
Specify the IP addresses
IP address:
see “Setting the Ethernet module parameters”, page 83
11
Do not click a COM port
12
Assign a name for this search
13
“Finish”
The Ethernet interface module (
see “Options for MCU control unit”, page 123
) must be
installed in the MCU (
see “Fitting the interface and I/O module (option)”, page 53
) and
configured (
see “Setting the Ethernet module parameters”, page 83
) to connect to the
measuring system via Ethernet.