O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
3
ASSEMBLY AND INSTALLATION
3.2.2
Fitting the MCU control unit
Fit the MCU control unit in a protected location that is easily accessible (
). Observe the following points during fitting:
●
Maintain the ambient temperature according to the Technical Data; take possible
radiant heat into consideration (shield when necessary).
●
Protect against direct sunlight.
●
Whenever possible, choose an assembly location with minimum vibrations; dampen any
vibrations when necessary.
●
Provide sufficient clearance for lines and opening the door.
Assembly dimensions
Fig. 20: MCU assembly dimensions
c
f
e
> 2
50
d
b
M8
Clearance to open door
Clearance for cables
a
Dimen-
sion
Control unit type
MCU-N
MCU-P
a
160
260
b
320
420
c
210
300
d
340
440
e
125
220
f
> 350
> 540
MCU-N:
Control unit without purge air supply
MCU-P:
Control unit with purge air supply