23
8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
PRODUCT DESCRIPTION
2
2.2.4
MCU control unit
Control unit MCU has the following functions:
●
Control of the data traffic and processing of the sender/receiver unit data connected via
the RS485 interface
●
Signal output via analog output (measured value) and relay outputs (device status)
●
Signal input via analog and digital inputs
●
Power supply for the connected measuring unit via 24 V switch-mode power supply with
wide range input
●
Communication with higher level control systems via optional modules
Equipment and device parameters can be set easily and conveniently via a USB interface
using a PC and a user-friendly operating program. The parameters are stored reliably even
in the case of a power failure.
Control unit MCU has a sheet steel enclosure as standard.
2.2.4.1
Standard interfaces
Analog output
Analog inputs
Relay outputs
Digital inputs
Communication
3 outputs 0/2/4...22 mA
(electrically isolated, active) to
output:
●
Dust concentration ext. a.c.
●
Dust concentration ext. s.c.
●
Dust concentration SI. a.c.
●
Dust concentration SI. s.c.
●
Scattered light
●
Opacity
●
Extinction
●
Transmission
●
Relative opacity
Resolution 10 bits
2 inputs 0...20 mA
(standard; without
electric isolation);
resolution 10 bits
5 changeover
contacts (48 V, 1 A)
to output status
signals:
●
Operation/failure
●
Maintenance
●
Function check
●
Maintenance
request
●
Limit value
4 inputs to connect
potential-free
contacts (e.g., to con-
nect a maintenance
switch or trigger a
function check)
●
USB 1.1 and
RS232 (on termi-
nals) for measured
value inquiries, set-
ting parameters
and software
updates.
●
RS485 for sensor
connection