TROUBLESHOOTING
Malfunctions listed below can probably be cleared onsite.
Message
Significance
Possible cause
Action
AO configuration
the number of available and
configured analog outputs is not
identical.
●
No parameters set for
AO
●
Connection error
●
Module failure
▸
Check configuration (
ting the analog outputs parame-
ters”, page 71
▸
Contact SICK Customer
Service.
AI configuration
Number of available and
configured analog inputs not
identical.
●
No parameters set for AI
●
Connection error
●
Module failure
▸
Check configuration (
ting the analog inputs parame-
ters”, page 74
▸
Contact SICK Customer
Service.
Interface Module
No communication via Interface
module
●
No parameters set for
module
●
Connection error
●
Module failure
▸
Check configuration (
ting the Ethernet module
parameters”, page 83
).
▸
Contact SICK Customer
Service.
No sensor found
Sender/receiver unit was not
recognized
●
Communication prob-
lems on RS485 line
●
Supply voltage problems
▸
Check system settings.
▸
Check the connection line
▸
Check voltage supply.
▸
Contact SICK Customer
Service.
Variant configuration
error
MCU setting does not match the
connected sensor
Sensor type has been
changed
▸
Correct application settings
(
sender/receiver unit”,
page 68
).
Testmode enabled
MCU in “Test” mode.
▸
Deactivate “System Test” mode
(“Maintenance” directory)