6
TROUBLESHOOTING
Message
Significance
Possible cause
Action
Reflector communication
No connection between
sender/receiver unit and
reflector
●
Connection line not connected
or not connected correctly
●
Defective connection line
●
Defective reflector/scattered
light receiver
●
RS485 interface of the sender
unit defective
▸
Check the connection line
▸
Contact SICK Customer
Service.
LED monitor overflow
Overmodulation of the
monitor channel during
scaling
Incorrect alignment of the opti-
cal axes of sender/receiver unit
and reflector
▸
Check/correct alignment.
▸
Repeat scaling
Q1-4 overflow
Group signal of quad-
rant measurement too
high
●
Measuring system not scaled
●
Changed alignment of optical
axes
●
Reduced active measuring
path
▸
Scale measuring system.
▸
Check/correct alignment.
▸
Contact SICK Customer
Service.
Set reference
Scaling not possible
Measurement or monitor signal
too low (contamination, incorrect
alignment)
▸
Check/correct alignment.
▸
Clean the optical surfaces
(
Contamination
Contamination value is
higher than the permit-
ted limit value (
●
Deposits on the optical inter-
faces
●
Unclean purge air
▸
Clean the optical surfaces
(
▸
Check purge air filter (
▸
Contact SICK Customer
Service.
Power supply (24 V) < 18 V Supply voltage too low
●
Line provided by customer
does not match the specifica-
tion (
●
Voltage loss on the connection
line (core cross-section too low
in relation to the line length)
▸
Check the connection line
▸
Contact SICK Customer
Service.
Power supply (24 V) < 19 V