"Steam sterilizer Laboklav 55 - 195
"
4.4.10. End of program
After successfully running the selected program, it is ended.
After the end of a program, the control waits for the operator to acknowledge the program!
This state is indicated by a flashing green display and a short acoustic signal. Acknowledgment is
done by pressing the
STOP button (11).
After a further 3 seconds the door is automatically unlocked.
Hot steam can escape when the chamber is opened!
Wait until the steam has completely escaped and the chamber
cools down before you remove the sterilized items.
After completing the program, please note the following when removing the sterile goods:
•
Never open the door with force.
The autoclave could be damaged and hot steam could escape
•
Do not touch the sterile items, the kettle or the door with unprotected hands.
The parts can be hot.