"Steam sterilizer Laboklav 55 - 195"
User manual
19
3.5. Programs for testing
The Bowie&Dick-Test is proofing the process for steam penetration of the sterilization good (mainly
for textiles or paper wrap). This test is implemented in vacuum option only. The test is normally for
medical use only! The test body for this test should be an usable test package only.
The vacuum test is implemented in vacuum option only. It can be used for testing the leakage rate of
the unit. For the normal use it is not necessary to run this test (implemented for service/maintenance).
The test should run if the chamber is cold! The chamber temperature should be not higher than 40°C
(otherwise it is impossible to start the test).
If the test result is “Incorrect” the unit should get a service /
maintenance! In that case you should not use the device until
qualified service stuff has checked the device!
3.6. Overview for predefined programs
All sterilization cycles are running automatically. The duration of a single cycle depends on the load,
kind of deaeration, number of vacuum fractions in fractionated pre-vacuum and the start conditions
(warm or cold), drying time, kind of cooling etc. Even so the type of sterilization goods and kind of
loading the good inside the chamber have an enormous effect on the cycle time. In case of a
validation you can define the goods and loads. When ever the same good / load with same program
and same parameters is started, the time depends mainly from start temperature. Basically, the
higher the mass of the load and the colder the items to be sterilized and the sterilization chamber, the
longer the time per cycle.
The table gives an overview for the different preselected programs:
(later changes are possible)
Prog.
Material
Deaeration
Steril.
time
Steril.
temp.
drying/coo
ling
P1 (B/M/MS/MSL)
Instr. ST
Instruments
Pressure purge:
3-5x 170 kPa / 120 kPa
15 min
134°C
No drying
P1 (V/MV/MSV/MSLV)
Instr. FV
Instruments
steam:
30/40/50kPa / 140kPa
15 min
134°C
5 min
P2 (B/M/MS/MSL)
Instr. ST
Instruments
Pressure purge:
3-5x 170 kPa / 120 kPa
15 min
134°C
No drying
P2 (V/MV/MSV/MSLV)
Instr. FV
plastics (e.g.
pipettes)
steam:
30/40/50kPa / 140kPa
25 min
134°C
40 min
P3 (B/M/MS/MSL)
Instr. ST
Plastics and gum
ware
Pressure purge:
3-5x 170 kPa / 120 kPa
20 min
121°C
No drying
P3 (V/MV/MSV/MSLV)
Instr. FV
Plastics and gum
ware
steam:
30/40/50kPa / 140kPa
20 min
121°C
15 min
P4 (B/M/MS/MSL)
Destruction ST
Waste / Trash
(solid mat. only!)
Pressure purge:
3-5x 170 kPa / 120 kPa
15 min
134°C
No drying
P4 (V/MV/MSV/MSLV)
Destruction FV
Waste / Trash
(solid mat. only!)
steam:
30/40/50kPa / 140kPa
15 min
134°C
No drying
P5 (B/M/MS/MSL)
Destruction ST
Waste / Trash
(solid mat. only!)
Pressure purge:
3-5x 170 kPa / 120 kPa
15 min
134°C
No drying
P5 (V/MV/MSV/MSLV)
Destruction FV
Waste / Trash
(solid mat. only!)
steam:
30/40/50kPa / 140kPa
15 min
134°C
2 min