"Steam sterilizer Laboklav 55 - 195"
User manual
43
8. Documentation of sterilization cycles
The device is prepared for 2 different types of process documentation. On the one hand, each
process can record the relevant process data directly on paper using a printer. At the same time, the
device contains a memory that stores between 50 and 1000 cycles depending on the number of
analog channels defined for documentation and the number of values to be saved per time unit. This
data can be recorded and archived at the same time as the current program using electronic
documentation software called
Sego Soft
. The cycles can be assigned accordingly using the date
and time. If the device control memory is full, the oldest data in the memory are automatically
overwritten.
With the
touch version
, the data is stored on an internal MicroSD card and can be called up at any
time via network or a connected USB stick (
do not leave the stick inserted, readout only takes
place directly after insertion or removal and re-insertion!
)
You can obtain separate instructions for the electronic documentation program on request from
customer service or in the customer center on the website:
www.shp-steriltechnik.de
The CBM-920 CITIZEN printer is a dot matrix printer. It is available as an integrated device which is
connected to the device via an RS 232 serial interface. The printer enables all relevant process
parameters to be printed out during or after the cycle.
The printer also enables the subsequent printing of process
data directly from the memory. To do this, select the Archive
functions submenu in the main menu.
The printer should have enough paper.
When the paper roll is used up, the printer switches off
automatically.
9. Decommissioning
If you want to take the device out of operation for a long time (e.g. due to a vacation or a planned
transport), proceed as follows:
•
The chamber door should not be completely closed, but only leaned against the locking
spindle so that the inside of the chamber can dry.
•
If there is a large amount of liquid in the chamber, it is advisable to remove it manually (always
pay attention to suitable protective clothing)
•
Empty the storage tank by using the drain hose on the quick coupling provided.
•
If necessary, empty the vacuum pump by using the drain plug on the pump head.
•
All supply lines such as cold water, demineralized water, compressed air and waste water
should be closed.
•
Disconnect the device from the mains.