44
User Manual
"Steam sterilizer Laboklav 55 - 195
"
10. Maintenance
The sterilizer should get regular cleaning, maintenance and service. Some parts are regular to be
changed completely to protect the device against damage or mistakes in sterilization cycle. The
simple cleaning and maintenance activities can be done by the operator without problems.
Special services can be done by specially trained service stuff only! All inspection activities acc. to
pressure vessel regulations / local regulations for pressure vessels and electrical installations need
special trained service stuff! We recommend to order one regular safety inspection per year and to
connect this with a regular maintenance for the vessel, pressure parts and electrical installations.
Your distributer is authorized to recommend a trained service partner.
We recommend you to use a device book that helps to
document all cleaning, maintenance and service repair
activities at the device.
Maintenance or repair activities that need to open the
housing are not allowed to be done by untrained personnel
stuff!
For maintenance or repair activities that need to open the
housing the electrical power supply must be disconnected!
Inside the housing dangerous electrical voltage can kill or
hurt!
After work at electrical installations some electrical test are
necessary to do. Please respect the local rules and
regulations!