5. Sterilization cycle progress
The program cycle is running fully automatically. The display shows the actual program cycle and
gives information what is the actual situation in the running program phase.
The successful finish of sterilization cycle will be displayed. In case of an incorrect cycle additionally
sounds an acoustic signal.
In the following the typical display are described:
Touchscreen (optional):
The steam sterilizer is switched on, program P1 is
activated but not started. The device is ready for
start. If the door is closed, the device is preheating
the steam generator automatically.
The display shows program no. P1. Here the
program type is shown, if special program name
was given it will replace the program type. The main
program parameters are shown for fast identification
of the program cycle. In the picture the display
shows Program type Instruments, with deaeration
fractionated prevacuum starting with vacuum
followed by steam injection 2 times. Sterilization is
programmed for 134°C, 4 min. sterilization time.
Drying function is not programmed. In the sensor
window the main sensor values are shown.
Touchscreen (optional)
Example for an error: The program P1 was started
but lid was not completely closed. So the device
generates an error message. The error message
can be cleared by pressing the Escape button if the
cause of the error was cleared.