"Steam sterilizer Laboklav 55 - 195"
User manual
7
The steam sterilizer line Laboklav 55 – 195 includes the following additional advantages:
-
Automatic preheating to reach defined condition in the start situation in each program separately
to define. This function is especially developed for fractionated prevacuum process to reduce the
condensate in the chamber and on the sterilization goods.
-
Fully automatic deaeration of the sterilization goods by fractionated prevacuum (in vacuum option
only) or pressure purge process for solids and gravity process for liquid goods.
-
Use of the principle of hot stone for the steam production reduces the use of water and decreases
the heating time of the device. That makes the process much more economic than the typical
steam generation process in a classic steam generator.
-
Cooling process using the feed water stored in an included feed water tank (in all units with
included fast cooling option). That reduces the need of expensive demineralized water and saves
heating energy in the feed water. Additionally it is degassing the feed water for the next cycle so it
increases the quality of the complete sterilization process.
-
Protection of the drain on house installation side by using normal tape water mixing to the steam
outlet. That process works temperature controlled by a PT 100 in the outlet installation. The
temperature is programmable for opening and closing the mixing valve.
-
Automatic cleaning function for the steam generator in running in each cycle. It protects the unit
against heater damage forced by bad water quality.
-
Microprocessor controlled process for fully automatic use.
Total view of the steam sterilizer Laboklav xxx
Figure 1: Total view of Laboklav xxx
The Laboklav should be operated while the normal laboratory working hours. If the device is not in
use it should be switched off by the main power plug (5). Over night and on the weekends the central
power plug (1) should be switched off. If that central power plug is missing, please use the main
breaker below the housing (4) on the right side of the unit.
Emergency Switch off
In case of a fatal switch off the device by turning the central
main power switch (1) or disconnect the sterilizer from power
supply by using the power plug (2).