"Steam sterilizer Laboklav 55 - 195"
User manual
27
4.4.5. Close the door/ lid
To close the device, press the door/ lid carefully against the locking spindle with your left hand while
pressing the
cursor down key (5)
with your right hand. Keep this button pressed until the closing
process is completed. You can see this on the display by the lock symbol, which now signals a closed
device, and the warning tone can no longer be heard. The closing process runs in 3 steps!
Keep the
cursor down key pressed all the time!
If you release the button without closing the lid completely,
you have to open the lid again before you can close it again (dead man circuit as a safety element).
If you or others are in danger of pinching your fingers or anything else, release the cursor down key!
The closing process stops automatically. In an emergency, switch off the device completely using the
power switch on the right of the control panel!
When closing, make sure that no parts are trapped between
the door and the seal!
Do not open and close the door by force, as this can damage
the door mechanism!
Safe operation cannot be guaranteed if damaged.
4.4.6. Choose a program
Program change
Activating a program is done by pressing
program button (9).
It opens the program menu. With the
button up (4) and down button (5)
the right program is chosen by pressing the
enter button (7)
. All
programs which are marked with a key symbol needs entering a code before activation:
The cursor buttons navigate the cursor. Up and down button changes the value. The enter button (7)
must be pressed to confirm the code.
2. Display level
In second display level the actual value of all installed sensors is shown.
You can switch to a second display level with the
display key (3).
The values of other sensors are
displayed here.
The symbols show the following sensors:
•
Tk – chamber temperature,
•
Tr – reference temperature,
•
Pk – chamber pressure,
•
Tg – temperature steam generator,
•
Tf – temperature in filter cartridge (Option FA only),
•
To – temperature in the steam / condensate outlet,
•
Tm – temperature in the outlet of heating / cooling jacket (not in all options)
•
Pg – pressure of steam generator
•
Wq – water conductance