TOC EVOLUTION:
USER &
MAINTENANCE MANUAL
Page:
58/74
Instructions No.: 10844
This document is the property of SERES and cannot be reproduced without authorization
11.2.2
Manual measure TOC
The "Manual measure (or Control measurement)" programme is used to check or determine the
concentration of a standard solution or sample.
It is often used in cases where the chemical method requires some measurements before giving a
significant result. One or several manual steps are then carried out before a calibration to obtain a correct
calibration coefficient.
From the main screen:
Make sure to stop the cycle in progress (from the main screen, press
).
Enter the authorised user code.
Access the "Tools" tab:
Push on Maintenance
;
Start the “TOC manual measure” by pressing
.
At the end of the cycle, a window displays the concentration measured under the name “manual
measurement”.
It’s the same thing for “TIC manual measure”.