To set the JBUS slave number, go to the "Tools" tab, "Options"menu.
The slave address is:
55
(by default).
JBUS by TCP/IP
When the client wants to connect via Ethernet, the analyzer is equipped with a converter at the top of the
device.
The connection is made via the Ethernet connector available on the right side of the analyzer on the
outside.
Simply mount the Ethernet cable in the connector and plug the connector.
For this communication, it is necessary to know the addressing mode desired by the customer. This
is important because the module must be configured correctly.
Two addressing modes:
- Automatic by a DHCP system which assigns an IP address to the analyzer
- With a fixed IP address provided by the client and which the manufacturer must know to configure
The Port number to configure is: 502.