TOC EVOLUTION:
USER &
MAINTENANCE MANUAL
Page:
39/74
Instructions No.: 10844
This document is the property of SERES and cannot be reproduced without authorization
o
Test the relevant output for the 3 output levels: 0%, 50% and 100%, respectively, 4, 12
and 20 mA. To do this, press the corresponding button.
8.3.3.1.6
Client input tab
This window allows you to visualise the state of the inputs.
This window is empty if no input has been requested by the customer.
8.3.3.2
Options (Setpoints, date/time, etc.)
It is possible to modify the setpoints and to set the time and date.
From the main screen:
Select the
“
Tools
”
tab.
Enter the authorised user code and validate.
Access the
“
Options
”
menu.
In the Options tab, it is possible to modify the temperature, flowrate setpoints, etc. It is strongly
recommended not to change these parameter settings without contacting SERES.