7 Technical specifications
7.1
The environment
Recommendations for the facility:
Room temperature between 5 and 40°C maximum
7.2 Characteristics
7.2.1
Measuring ranges
3 measuring ranges are available in two versions:
- Range 1: Dynamic ranges 0-2 ppm (Dual channel configuration possible)
- Range 2: Dynamic ranges 0-10 ppm (Dual channel configuration possible)
- Range 3: Dynamic ranges 0-100 ppm (Dual channel configuration possible)
7.2.2
Measuring rate
5 to 10 min according to the range but it also depends on the nature of the sample and on the type of
measurement.
The spacing between the measurements can be adjusted
.
7.2.3
Standard dimensions
Casing:
l x h x d: 993x600x422 mm
Volume: 0.25m
3
Weight
: 80kg
Protection:
IP55
If there are additional options, please consult the specific dimensional drawing
.
7.2.4
Electrical characteristics
Consumption:
300 VA maximum
Input supply:
240 V, 50/60 Hz or 110V (configuration)
7.2.5
Sample inlet
Temperature
: min: 5°C max: 50°C
Clean - weakly charged sample:
Flow rate:
50 l/h (if the fast sampling loop option is available)
Pressure:
Room pressure (in a tank)
-
3 bars maximum if a fast sampling loop is available
Charged sample:
Sampling & Treatment
: Particles stored up to 100
μm
(please consult us)
–
Flowrate for
Sampling/treatment system (100 to 500l/h
–
1bar)