TOC EVOLUTION:
USER &
MAINTENANCE MANUAL
Page:
53/74
Instructions No.: 10844
This document is the property of SERES and cannot be reproduced without authorization
10.7 Timing of sample purging line (Option on specific customer request)
If this option is requested, it is possible to put a time before starting the analysis. This time
corresponds to the purge time of the sample line in the event of distance between the analyzer and the
customer's sampling point.
This time can be modified in the “Option” menu (See paragraph 8.3.3.2) on the HMI and must be set
in seconds.
Before each measurement cycle, the analyzer will wait this time before taking sample and measuring.
This time is called (in the Option menu): “Timing of sample purging line”.
10.8 Launch of the continuous cycles
In order to start the continuous analysis, press button
in the main window.
10.9 Visualisation and recovery of the data
The measurements are stored in the screen memory. They can be transferred to a USB key or
displayed on a graph.
Every time the analyser is switched off, the graph and history empty, however the data remain stored
in the analyser.
To recover the data:
Open the door of the analyser, insert the USB key into the port:
From the main screen:
Press the
“
Tools
”
tab.
Enter the authorised user code
Go to the "File Copy Log" menu.
Press the validation symbol of the window that opens.
To read the data on a computer: Connect the key to the computer.
Select the "logs" file.
The uploaded data are stored on a daily basis (YYYY_MM_DD)