Sequencer COMPACT
Technical brochure
FI 72.0467.0118E
Page 40 / 42
WARNING
-avoid input of dusty air in the pressure connections.
-respect device characteristics (overvoltage, overpressure …)
- For the connection of the 4-20mA and the link towards CDM sensor, it is recommended to use an shielded cable and
not follow the power voltage cable.
-for specific exploitation conditions: consult us
-interventions must always be done by a qualified person
CALIBRATION
When the device leaves our workshop, it is adjusted at 0 value for the zero pressure value and for the maxi pressure
value of the range.
Adjustments are done when the device is stabilized and at an ambient temperature of 20°C.
It is possible to adjust the zero pressure:
-disconnect the 2 pressure connections
-access to adjustment via page « pressure switch parameters »
-connect the 2 pressure connections