![Sartorius LMA110S Operating Instructions Manual Download Page 76](http://html1.mh-extra.com/html/sartorius/lma110s/lma110s_operating-instructions-manual_1199621076.webp)
76
To change the temperature adjustment schedule:
§
From the Service menu, press the
3
key to display
the Adjust Temperature screen.
§
Press the 3 for Adjust schedule. The screen will
change to the Schedule temp adjust screen.
§
Select the desired period of time. When this period
expires, a warning will appear in the warning box
indicating that a new verification or temperature
adjustment should be performed.
To enter a reference fixture number:
§
From the Adjust Temperature screen, press the
4
for Reference fixture. The screen will change to
have you enter the reference fixture number that is
marked on your temperature bowl. The number will
then appear on both the temperature adjustment or
verification reports.
1 ) O f f
2 ) 1 d a y
3 ) 7 d a y s
4 ) 3 0 d a y s
5 ) 9 0 d a y s
6 ) 1 8 0 d a y s
7 ) 3 6 5 d a y s
S c h e d u l e c a l i b r a t i o n
C a l i b r a t i o n s c h e d u l e : O f f
0 5 / 0 1 / 2 0 1 1
0 2 : 5 0 P M