Start Delay
The start delay is a time in seconds that can be set
to ensure that a good initial weight is taken at the
beginning of the test.
To set or change start delay:
§
Press the
5
key from the Program Menu to display
the Start delay screen. The current start delay is
shown.
§
Use the numeric keys to type in the desired %IW,
then press the
Enter
key.
§
Note the mode: Mark 2 or LMA110S. If in the Mark 2
mode, start delay will simulate the previous Mark 2
analyzer. Press the
Mode
soft key to toggle between
Mark 2 and LMA110S.
§
Press the
Enter
key again or the
Arrow
soft key to
return to the Program Menu.
Standby Temperature
The standby temperature is the temperature that
the analyzer chamber will be controlled to at the
end of a test or prior to a test when a new program
is recalled.
To set or change standby temperature:
§
Press the
6
key from the Program Menu to display
the standby temperature screen. The current standby
temperature is shown.
§
Use the numeric keys to type in the desired standby
temperature, press the
Enter
key.
§
When all the parameters for slope have been
changed, press the
Enter
key again or
Arrow
key to
return to the Program Menu.
36
E n t e r s t a n d b y t e m p : 6 0 C
M a x i m u m t e m p e r a t u r e 1 6 5 C
S t a n d b y T e m p e r a t u r e
D e f a u l t
| C
Clear
0 5 / 0 1 / 2 0 1 1 0 2 : 3 1 P M