The automatic temperature adjustment report is
printed after the adjustment process.
To verify an adjustment temperature:
§
From the Service menu, press the
3
key to display
the Adjust Temperature screen.
§
Press the
2
for Verify temperature. The screen will
change to the Verify temperature screen.
§
The default verifies temperature is 130°C. Press the
“Temp” soft key to begin. The screen will change to
the Temp verify in process screen and show the
selected temperature. After 20 minutes, the screen
will change to the Temp verify done screen and will
display the expected selected temperature and the
actual temperature that was measured by the
temperature bowl (“fixture”).
§
Press the
soft key to print a temperature
verification report (optional).
§
If the agreement of the selected temperature and
the actual temperature are not within 2 degrees for
any of the selected temperatures, it is recommended
that you perform the temperature calibration and
adjustment procedure.
Automatic Temperature Adjustment Report
- - - - - - - - - - - - - - - - - - - - - - - - - -
Fixture number 1234
Temperature 80C: 82C
Temperature 150: 150C
Mark 3 LTE Moisture Analyzer
Module SN: M3085010B
Calibration passed 05/10/11 02:05 PM
Signature:
__________________________
P r e a d j u s t 8 0 : 7 9
A d j u s t e d 8 0 : 8 0
P r e a d j u s t 1 5 0 : 1 4 4
A d j u s t e d 1 5 0 : 1 5 0
T e m p a d j u s t d o n e
0 5 / 0 1 / 2 0 1 1 2 0 2 : 4 3 P M
A d -