Functions available from the Standby screen:
§
Press
Start|Stop
key to initiate a test
§
Press
Recall
soft key to change the program to
be used for testing a sample
§
Press
User
soft key to select an operator name
(feature must be turned on in Setup)
§
Press
Setup
soft key to enter the Setup menu in
order to make all parameter changes
§
Press
Graph
soft key to display a graph
§
Press
Weigh
soft key to use as a balance
Standby-Temperature
!
with a New Program
When creating a new program, the standby
temperature setting from the previous drying
program is used automatically by the analyzer.
Example:
You have created a new program for drying
a temperature-sensitive sample at low tempera-
ture. The analyzer uses the standby tempera-
ture, e.g. 165°C, from the previous program.
You must call up the standby temperature
parameter and change the temperature value.
If you forget performing this setting, your ana-
lyzer is always too hot before starting the test.
22