4. At the “Add sample to pan” prompt, open the heater
hood and begin adding sample to the pan. A bar
graph on the display will guide the operator to place
the correct amount of sample as set by the ideal
weight parameter of the drying program. Continue
adding sample until the beeper sounds, indicating
that the sample amount is within 10% of the ideal
weight. The exact weight is also displayed on the
screen.
5. At the “Close heater hood” prompt, close the heater
hood. The balance will capture the initial weight of
the sample “Sampling” and then show that the test
is in progress. The display will change to show cur-
rent information including the elapsed time and the
current percentage weight loss in the units as speci-
fied in the drying program.
Note
To stop the test in progress, press the
Start|Stop
key.
The optional printer will begin to print as the test
begins. Printout information can be selected in
Setup menu under Printer.
At the end of the test the display will show “Test
Done.” The result will stay on the screen as well as
be printed. The analyzer will enter the Standby
mode until the next test is initiated.
26
D r y M i l k P o w d e r
T e s t i n P r o g r e s s
0.01
% Moisture
E l a p s e d t i m e : 0 0 . 1 0
Graph
P r e s s S T O P k e y t o s t o p t e s t
Temp 1
Current
105
71
0 5 / 0 1 / 2 0 1 1
0 2 : 1 8 P M
P r e s s S T A R T k e y t o b e g i n t e s t
D r y M i l k P o w d e r
T e s t D o n e
3.47
% Moisture
R e s u l t t i m e : 7 . 0 m i n
I D : 1 2 6 3 9 9 . 1
Target: Pass
Recall
Setup
Graph
User
W a r n i n g : S a m p l e o n p a n
Standby
Current
105
105
0 5 / 0 1 / 2 0 1 1 0 2 : 1 8 P M