Mark 3 LTE Program Download Format
Introduction:
This document details the program output format for the Mark 3 LTE
Moisture Analyzer. The analyzer has the capability to transfer program settings to another
Mark LTE 3 or a PC. The analyzer can also be downloaded with settings from a PC. This
text format is used to transfer all the information necessary to control the test conditions
of a moisture analysis test.
Field
Description
1
Program Number (#XXX)
2
Program Name (15 characters max.)
3
Units (0=Moisture, 1=Solids, 2=Volatiles, 3=Moisture dry,
4=PPM Moisture, 5=Weight, 6=mg/L)
4
Slope Time (X.X minutes)
5
Slope Percentage (X.XXX)
6
Slope Calculation (0=Actual, 1=Calc 1, 2=Calc 2, 3=Calc 3)
7
Standby Temperature (165C max)
8
Equilibration Time (X minutes)
9
Ideal Weight (XX.X grams)
10
Ideal Weight % (XX.X)
11
Ideal Weight Lock (0=OFF, 1=ON)
12
Target Limit: Upper (X.XXX %)
13
Target Limit: Lower (X.XXX %)
14
Target Limit: Usage (0=OFF, 1=ON)
15
Test Resolution (2=X.XX%, 3=X.XXX%)
16
Scale Factor (X.XXX) Default=1.000
17
Scale Offset (X.XXX) Default=0.000
18
Temperature 1 (XXX)
19
Ramp 1 (0=Standard)
20
Time 1 (XX.X)
21
Temperature 2 (XXX)
22
Ramp 2 (0=Standard)
23
Time 2 (XX.X)
24
Test Mode (0=Standard, 1=Concentration, 2=Fat)
25
Start Delay (seconds)
Output Example:
# 1,Default,0,1.0,0.050,0,60,0,5.0,10.0,0,100.00,0.000,0,3,1.000,0.000,105,0,0.0,0,0,
0.00,128/r/n