![Sartorius LMA110S Operating Instructions Manual Download Page 30](http://html1.mh-extra.com/html/sartorius/lma110s/lma110s_operating-instructions-manual_1199621030.webp)
Table: Program Settings
Screen
Parameter
Variables
Explanation
Units
Units
Moisture
Initial wt. – final wt. / Initial wt.
+
100
Solids
Final wt. / Initial wt.
+
100
Volatiles
Initial wt. – final wt. / Initial wt.
+
100
Moisture/dry
Initial wt. – final wt. / Final wt.
+
100
mg/l
Final wt.
+
1000 / volume in mL
Weight
Initial wt. – final wt.
Result display
Decimal place
0.01% or 0.001%
Factors
Scale factor
Factored result = (result
+
scale factor) + offset
Offset
Factored result = (result
+
scale factor) + offset
Temperature 1
Temperature
Maximum 210°C
Set temperature or Off
Time
10.0 – 99.9 minutes
Set time or Off
Temperature 2
Temperature
Maximum 210°C
Set temperature or Off
Time
20.0 – 99.9 minutes
Set time or Off
Slope
% Initial weight
0.01 – 9.99
Weight variable of endpoint criteria
Window of time
0.1 – 99.9 minutes
Time variable of endpoint criteria
Calculation
Actual slope
Slope criteria met for endpoint
Calc 1
Calculated endpoint
Calc 2
Calculated endpoint
Calc 3
Calculated endpoint
Start delay
Start delay
0 - 9 seconds
Delay before starting test
Mode
Mark 2
Simulates Mark 2 start delay
LMA110S
LMA110S start delay
Standby temp.
Standby temp.
Maximum 165°C
Set temperature between tests
Ideal weight
Ideal weight
0.1 – 99.0 g
Set ideal sample weight for test
Percent
1 – 10%
Set tolerance of ideal weight
Ideal weight
Lock On or Off
Set ideal weight lock strict (On) or Off
Target limit
Target limit
Upper limit
Set upper limit of result tolerance
Lower limit
Set lower limit of result tolerance
Turn limits
Off
Turns target limit feature Off
Mode
Mode
Standard
Standard test mode
Concentration
Corrects a result for sample volume