Fig.2.11 Measurement interface in case of connection with a peak power sensor
Table 2.7 Description of operation interface
Graphic
Number
Information Content Meaning
1
Power display
It displays the current measured power value.
2
Power unit
It displays the current power unit including dBm and W.
3
Now gate
It displays the gate shown by the current measurement result, A1 for Gate 1
of Channel A.
4
Numeric
measurement type
It displays the measurement type of the gate shown by the current
measurement result, including “AVG PWR”, “Peak PWR”, “PAR”, “Min
PWR” and “Max/min”, “AVG” indicates that the numeric measurement type
is “AVG PWR”.
5
FDO table
It displays the called FDO table shown by the current measurement result.
There are ten optional FDO tables. The FDO table is a frequency factor
correction table defined by the user.
(1) represents calling of the first FDO table.
6
Frequency value
It is the carrier frequency value of the current measured signal, which should
be manually set the user.
7
Trigger mode
It displays the trigger mode. It sets the trigger mode for the current
measurement, including “Free Run”, “Continue Trig” and “Single Trig”.
represents “Continue Trig”.
8
Offset state
There are two offset states. If the display offset is enabled, the "Disp Offset
On" will be displayed. If the channel offset is enabled, the “Channel Offset
ON” will be displayed. If both offsets are enabled, “Offsets ON” will be
displayed. The display offset is used for the current window; the channel offset
is used for the selected channel.
9
Instrument state
It displays the instrument is in local or remote control state currently.