Chapter 8: RESP Monitoring
User Manual
8-5
8.3 RESP Alarm Messages
a) Physiological alarms
The alarm is activated when the respiration rate exceeds the adjusted alarm limits.
b) Technical alarms
Physiological alarms
AUDIO
SOUND
VISUAL PROMPTS
SITUATION
ALARM
Activated
●
RESP value blinks
●
The alarm indicator flashes.
●
The alarm message is displayed in a
background corresponding to its level.
Respiration rate violates
adjusted high limit
RR HIGH
Activated
●
RESP value blinks
●
The alarm indicator flashes.
●
The alarm message is displayed in a
background corresponding to its level.
Respiration rate violates
adjusted low limit
RR LOW
Activated
● The alarm indicator flashes.
●"RESP APNEA" is
displayed in red background.
Non-respiration condition
overruns adjusted time
APNEA
RESP ALARMS
Alarm
Cause
Solution
Explanation
RESP CHECK
LEADS
The RESP leads are not properly
connected.
Make sure that all
electrodes, lead are
properly connected
Alarm level 3- the message is
displayed in cyan background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault. The alarm is activated when
RR ALARM is "ON".