TREND
User Manual
17-4
■
To change the display scale:
Pick "SCALE" (the second left item) to adjust the Y-axis scale and thus change the trend curve
in proportion.
PARAMETER
SCALE1
SCALE2
SCALE3
SCALE4
SCALE5
SCALE6
MIN
MAX
MIN
MIN
MIN
MAX
MIN
MAX
MIN
MAX
MIN
MAX
HR
0
60
0
120
0
240
-
-
-
-
-
-
SpO2
80
100
60
100
0
100
-
-
-
-
-
-
T1/ T2
30
42
24
48
0
48
-
-
-
-
-
-
IBP1/IBP2/
IBP3/IBP4
-20
50
-20
100
-20
200
-50
300
50
250
-
-
RESP
0
60
0
120
0
240
-
-
-
-
-
-
AWRR
0
60
0
120
0
240
-
-
-
-
-
-
Co2/ EtCo2/
FiCo2
0
50
0
100
-
-
-
-
-
-
-
-
O2/ EtO2/ FiO2
0
50
0
100
-
-
-
-
-
-
-
-
N2O/ EtN2O/
FiN2O
0
50
0
100
-
-
-
-
-
-
-
-
AA/ EtAA/
FiAA
0
1.0
0
2.0
0
3.0
0
5.0
0
10.0
0
20.0
PVCs
0
20
0
50
0
100
-
-
-
-
-
-
ST
-0.2
0.2
-0.5
0.5
-1
+1
-2
2
-
-
-
-
AFIB
0
1
-
-
-
-
-
-
-
-
-
-
PR
0
60
0
120
0
240
-
-
-
-
-
-
PI
0
20
0
10
0
5
-
-
-
-
-
-
PVI
0
30
0
100
-
-
-
-
-
-
-
-
SpOC
0
36
6
20
-
-
-
-
-
-
-
-
SpCo
0
12
0
24
0
50
-
-
-
-
-
-
SpMet
0
6
0
20
-
-
-
-
-
-
-
-
SpHb
6
20
2
14
0
25
-
-
-
-
-
-
■
To select time interval of displaying numeric values:
Pick the third left item, available options are 15 min, 30 min and 45 min, 1, 2 and 4 hours.
■ To select time intervals of trend in
x-axis
Select
or
to change time interval in the X-axis and to adjust start time and end time. By
every click on these buttons, you can change the time interval of x-axis to the extent of the
specified time in the third left item.