Chapter 24: Care and Cleaning
User Manual
24-9
SAADAT Co.
Form No. : PL-F-24
PM Form (BEDSIDE)
State: City: Medical center: Ward:
Device model: Serial number: Installation date: Inspection date:
N/A
NOK
OK
Test and Inspection Item
No.
Check parameters of RESP window
RESP
10
Check TEMP probe
TEMP
11
Cleaning and disinfection according to the user manual
Check SpO2 probe (extension, if any)
SpO2
12
SpO2 window settings (Measurement mode and
sensitivity)
Cleaning and disinfection according to the user manual
Check NIBP cuff and hose
NIBP
13
NIBP window settings (Adult, Pediatric and Neonate
modes, measurement unit, Automatic mode)
Cleaning and disinfection according to the user manual
Flushing the tubing system and perform zeroing
IBP
14
Check transducer and accessories
IBP window settings (Measurement unit, filter, Auto
Scale and etc)
Cleaning and disinfection according to the user manual
Check CAPNO probe and ISA Sampling line
CAPNO
15
Check CAPNO probe and IRMA Adaptor
CAPNO window settings (Measurement unit,
COMPENSATE and etc)
Cleaning and disinfection according to the user manual