
Chapter 5: ECG Monitoring
User Manual
5-18
NOTE:
If HR SOURCE is set to any signal except ECG, beat symbol and sound will be
according to the selected signal.
NOTE:
If HR SOURCE is set to any module except ECG, HR will change to PR and its
colour will change corresponding to the selected module for HR SOURCE.
NOTE:
If “HR SOURCE” is set to any module and cable of the module is not connected to
the system, HR value will not be displayed
NOTE:
IBP3 and IBP4 can only be active in the ALBORZ B9 system.
NOTE:
Calculating HR from IBP signal is possible just from ART, PAP, RVP, LVP and
IBP labelled signal.
NOTE:
Calculating HR value from IBP signal is not possible in the following conditions and
the HR value will be displayed
"
---
"
:
-
“IBP1/IBP2 STATIC PRESSURE
”
message on the display
-
“IBP1/IBP2 SEARCH” message on the display
-
HR value less than 25
-
Selecting CVP, LAP and RAP labels.
NOTE:
HR value measurement range is 25~240 bpm, when the HR is calculated from IBP
signal