Chapter 24: Care and Cleaning
User Manual
24-6
The following table summarizes the methods of cleaning, disinfecting and sterilizing different
parts of the device:
Device parts
Single-
use
Cleaning
Disinfection
Sterilization
External surface of device
-
In-between patients
and as required wipe
gently using a moist
cloth and warm soapy
water or mild
detergent.
In-between patients and as
required use
■
Alcohol 70%
■
Isopropyl alcohol
■ N
-propanol
To avoid extended
damage to the equipment,
sterilization is not
recommended for this
monitor, related products,
accessories or supplies
unless otherwise indicated
in the Instructions for Use
that accompany the
accessories and supplies
or when stipulated as
necessary in the Hospital
Maintenance Schedule.
BFA module
disposable
electrodes
* Trolley/ Wall stand,
* Holders of accessory,
* Extension cables
-
In-between patients and as
required use
■
Isopropyl alcohol
Display screen
-
In-between patients
and as required:
Clean and soft cloth
with screen cleaner or
mild soapy water
Recorder (printhead)
-
as required:
1.Gently wipe around
the printhead using
cotton swabs
dampened with
alcohol.
2.After the alcohol has
completely been dried,
reload the paper and
close the recorder
door.
use as required
■
Isopropyl alcohol
ECG Accessory
disposable
electrodes
According to the instructions delivered with the reusable accessories
To clean, disinfect and sterilize reusable transducers, sensors, cables, leads, and so
forth, refer to the instructions delivered with the accessory.
SpO2 Accessory
disposable
sensor
NIBP Cuff
-
TEMP Accessory
-
IBP Accessory
disposable
transducers
and Domes
GAS Accessory
(Main-stream/Side-stream)
disposable
Airway
Adapter,
Nemoline
family
sampling
lines
CO Accessory
-